Development and Validation of an Analytical HPLC Method to Assess Chemical and Radiochemical Purity of [68Ga]Ga-NODAGA-Exendin-4 Produced by a Fully Automated Method

Author:

Migliari SilviaORCID,Sammartano AntoninoORCID,Boss Marti,Gotthardt Martin,Scarlattei Maura,Baldari Giorgio,Silva Claudia,Bonadonna Riccardo C.,Ruffini LiviaORCID

Abstract

Background: Glucagon-like peptide 1 receptor (GLP-1R) is preferentially expressed in pancreatic islets, especially in β-cells, and highly expressed in human insulinomas and gastrinomas. In recent years several GLP-1R–avid radioligands have been developed to image insulin-secreting tumors or to provide a tentative quantitative in vivo biomarker of pancreatic β-cell mass. Exendin-4, a 39-amino acid peptide with high binding affinity to GLP-1R, has been labeled with Ga-68 for imaging with positron emission tomography (PET). Preparation conditions may influence the quality and in vivo behavior of tracers. Starting from a published synthesis and quality controls (QCs) procedure, we have developed and validated a new rapid and simple UV-Radio-HPLC method to test the chemical and radiochemical purity of [68Ga]Ga-NODAGA-exendin-4, to be used in the clinical routine. Methods: Ga-68 was obtained from a 68Ge/68Ga Generator (GalliaPharma®) and purified using a cationic-exchange cartridge on an automated synthesis module (Scintomics GRP®). NODAGA-exendin-4 contained in the reactor (10 µg) was reconstituted with HEPES and ascorbic acid. The reaction mixture was incubated at 100 °C. The product was purified through HLB cartridge, diluted, and sterilized. To validate the proposed UV-Radio-HPLC method, a stepwise approach was used, as defined in the guidance document released by the International Conference on Harmonization of Technical Requirements of Pharmaceuticals for Human Use (ICH), adopted by the European Medicines Agency (CMP/ICH/381/95 2014). The assessed parameters are specificity, linearity, precision (repeatability), accuracy, and limit of quantification. Therefore, a range of concentrations of Ga-NODAGA-exendin-4, NODAGA-exendin-4 (5, 4, 3.125, 1.25, 1, and 0.75 μg/mL) and [68Ga]Ga-NODAGA-exendin-4 were analyzed. To validate the entire production process, three consecutive batches of [68Ga]Ga-NODAGA-exendin-4 were tested. Results: Excellent linearity was found between 5–0.75 μg/mL for both the analytes (NODAGA-exendin-4 and 68Ga-NODAGA-exendin-4), with a correlation coefficient (R2) for calibration curves equal to 0.999, average coefficients of variation (CV%) < 2% (0.45% and 0.39%) and average per cent deviation value of bias from 100%, of 0.06% and 0.04%, respectively. The calibration curve for the determination of [68Ga]Ga-NODAGA-exendin-4 was linear with a R2 of 0.993 and CV% < 2% (1.97%), in accordance to acceptance criteria. The intra-day and inter-day precision of our method was statistically confirmed using 10 μg of peptide. The mean radiochemical yield was 45 ± 2.4% in all the three validation batches of [68Ga]Ga-NODAGA-exendin-4. The radiochemical purity of [68Ga]Ga-NODAGA-exendin-4 was >95% (97.05%, 95.75% and 96.15%) in all the three batches. Conclusions: The developed UV-Radio-HPLC method to assess the radiochemical and chemical purity of [68Ga]Ga-NODAGA-exendin-4 is rapid, accurate and reproducible like its fully automated production. It allows the routine use of this PET tracer as a diagnostic tool for PET imaging of GLP-1R expression in vivo, ensuring patient safety.

Publisher

MDPI AG

Subject

Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3