Author:
Lan Guanyu,Li Xue,Jia Hongyu,Yu Xiaofeng,Wang Zhaohui,Yao Jijun,Mao Xuefei
Abstract
In this work, a direct solid sampling device based on modified graphite furnace electrothermal vaporization (GF-ETV) with inductively coupled plasma mass spectrometry (ICP-MS) was established for the simultaneous detection of trace selenium and cadmium in rice samples. A bypass gas was first designed in GF-ETV to improve the device’s analytical sensitivity and precision. The ashing and vaporization conditions, the flow rates of the Ar carrier and the bypass gases of ICP-MS were all investigated. Under the optimized conditions, the limits of detection (LODs) for Se and Cd were 0.5 μg kg−1 and 0.16 μg kg−1, respectively; the relative standard deviations (RSDs) of repeated measurements were within 8% (n = 6). The recoveries of Cd and Se in rice samples were in the range of 89–112% compared with the microwave digestion ICP-MS method, indicating good accuracy and precision for the simultaneous detection of Se and Cd in rice matrix. The whole analysis time is <3 min without the sample digestion process, fulfilling the fast detection of Se and Cd in rice samples to protect food safety.
Funder
National Natural Science Fund of China
Key Research and Development Project of Hainan Province
Central Public interest Scientific Institution Basal Research Fund
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Cited by
10 articles.
订阅此论文施引文献