Abstract
A measurement and speciation procedure for the determination of total mercury (HgTOT), inorganic mercury (HgIN), and methylmercury (CH3Hg) was developed and the applicability for on-site determination was demonstrated. A simple, portable sample pretreatment procedure was optimized to extract the analytes. Home-made columns, packed with a new sorbent material called CYXAD (CYPHOS 101 modified Amberlite XAD), were used to separate the two forms of the analyte. HgTOT and CH3Hg were determined by anodic stripping voltammetry (ASV), using a solid gold electrode (SGE). Two certified reference materials (BCR-463 Tuna Fish and Tuna Fish ERM-CE 464) and eight fresh fishes were analyzed. Then, the results that were obtained following the optimized portable procedure were compared with the concentrations obtained, using a direct mercury analyzer (DMA). This quantification, using the two techniques, demonstrated the good performance of the proposed method.
Funder
European Union's Horizon 2020 research and innovation programme under the Marie Skłodowska-Curie
Subject
Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science
Cited by
2 articles.
订阅此论文施引文献