Comparison of LC-MS3 and LC-MRM Methods for Quantifying Amantadine and Its Application in Therapeutic Amantadine Monitoring in Human Plasma

Author:

Sun Qiang,Cao Haiwei,Liu Yong,Li Yanyan,Huang Jing

Abstract

A simple sample preprocessing method was developed for the quantitative determination of amantadine (AMT) in human plasma by liquid chromatography-tandem mass spectrometry cubed (LC-MS3). The LC-MS3 system comprised a Shimadzu Exion LC-20AD HPLC pump coupled with a QTRAP 5500 mass spectrometer. First, the plasma samples were pretreated using acetonitrile as the extracting solution to precipitate protein. Next, amantadine and amantadine-d15 (AMT-d15) were separated on an Agilent Poroshell 120 SB-C18 column (4.6 mm × 50 mm, 2.7 μm) using isocratic elution with solvent A (70% 0.1% formic acid) and solvent B (30% acetonitrile) at a flow rate of 0.8 mL/min. The total run time for each sample was 3 min. The system used triple-stage fragmentation transitions at m/z 152.2→135.3→107.4 for AMT quantification in the positive ion mode and m/z 167.0→150.3→118.1 for AMT-d15 quantification. The LC-MS3 assay was linear (r > 0.995) with a concentration range of 50–1500 ng/mL. The lower limit of quantification (LLOQ) was 50 ng/mL, and the intra-day and inter-day accuracies and precisions were less than 8.0% at all concentrations. In addition, the recoveries and matrix effect for AMT in human plasma were within acceptable limits. In terms of stability, AMT had no significant degradation under all conditions. All the results met the requirements of the guidelines of the Food and Drug Administration (FDA) for biological method validation. The novelty of the MS3 assay was that it presented a methodology with higher selectivity and sensitivity. This method was successfully applied to 44 human plasma samples, and the obtained quantitative results were compared with another liquid chromatography-multiple reaction monitoring (LC-MRM) method. The Passing-Bablok regression coefficients and Bland-Altman plot revealed no difference between the LC-MS3 and LC-MRM methods, implying that the developed LC-MS3 method is a reliable and accurate assay for AMT determination in human plasma. These results are also a proof of concept for determining chemicals in biological samples by the LC-MS3 strategy.

Funder

First Hospital of Jilin University

Project of Jilin Provincial Department of science and technology

Publisher

MDPI AG

Subject

Chemistry (miscellaneous),Analytical Chemistry,Organic Chemistry,Physical and Theoretical Chemistry,Molecular Medicine,Drug Discovery,Pharmaceutical Science

Reference28 articles.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3