Author:
He Quanguo,Tian Yaling,Wu Yiyong,Liu Jun,Li Guangli,Deng Peihong,Chen Dongchu
Abstract
4-nitrophenol (4-NP) is a hazardous waste and a priority toxic pollutant identified by US Environmental Protection Agency (EPA). Hence, in this paper, a voltammetric sensor was proposed for the direct and sensitive detection of 4-nitrophenol (4-NP) at nanomolar level in complex matrices by using graphene and acetylene black paste hybridized electrode (GR/ABPE). Under optimal conditions, the calibration curve demonstrates a linear relationship for 4-NP in the range from 20 nM to 8.0 μM and 8.0 μM to 0.1 mM separately with the detection limit of 8.0 nM. In addition to it, the performance of the GR/ABPE in practical applications was evaluated by detecting 4-NP in various water samples, and satisfactory recoveries were realized. Therefore, GR/ABPE may have a great potential application for facile and sensitive detection of 4-NP in complex matrices at nanomolar level.
Subject
General Materials Science,General Chemical Engineering
Cited by
120 articles.
订阅此论文施引文献