Abstract
In this work, we developed a novel all-3D-printed device for the simple determination of quetiapine fumarate (QF) via voltammetric mode. The device was printed through a one-step process by a dual-extruder 3D printer and it features three thermoplastic electrodes (printed from a carbon black-loaded polylactic acid (PLA)) and an electrode holder printed from a non-conductive PLA filament. The integrated 3D-printed device can be printed on-field and it qualifies as a ready-to-use sensor, since it does not require any post-treatment (i.e., modification or activation) before use. The electrochemical parameters, which affect the performance of the sensor in QF determination, were optimized and, under the selected conditions, the quantification of QF was carried out in the concentration range of 5 × 10−7–80 × 10−7 mol × L−1. The limit of detection was 2 × 10−9 mol × L−1, which is lower than that of existing electrochemical QF sensors. The within-device and between-device reproducibility was 4.3% and 6.2% (at 50 × 10−7 mol × L−1 QF level), respectively, demonstrating the satisfactory operational and fabrication reproducibility of the device. Finally, the device was successfully applied for the determination of QF in pharmaceutical tablets and in human urine, justifying its suitability for routine and on-site analysis.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献