Affiliation:
1. Istituto Nazionale di Ricerca Metrologica (INRIM), Strada delle Cacce 91, 10135 Torino, Italy
Abstract
A compact thermometer for air temperature based on the measurement of the speed of sound was developed at INRIM. This paper focuses on the comparison of this instrument with platinum resistance thermometers in a climatic chamber over a temperature range (−30 ÷ +55) °C, relative humidity (10 ÷ 90)%rh, and irradiation (>1 kW/m2) values similar to those of surface atmospheric conditions. Overall uncertainty values of 0.2 °C over the range from −30 °C to +30 °C, and from 0.6 °C to +55 °C, were found. Moreover, the instrument proved to be immune to irradiation errors and free from the need for temperature calibration.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献