Scattered Radiation Distribution Utilizing Three Different Cone-Beam Computed Tomography Devices for Maxillofacial Diagnostics: A Research Study

Author:

Petsaros Sotirios1,Chatzipetros Emmanouil1ORCID,Donta Catherine1,Karaiskos Pantelis2,Boziari Argiro3,Papadakis Evangelos1,Angelopoulos Christos1

Affiliation:

1. Department of Oral Diagnosis and Radiology, Faculty of Dentistry, National and Kapodistrian University of Athens, 2 Thivon Street, Goudi, 11527 Athens, Greece

2. Medical Physics Laboratory, Faculty of Medicine, National and Kapodistrian University of Athens, 75 Mikras Asias Street, Goudi, 11527 Athens, Greece

3. Greek Atomic Energy Commission, Agia Paraskevi, 15310 Attiki, Greece

Abstract

This study aimed to estimate scattered radiation and its spatial distribution around three cone-beam computed tomography (CBCT) devices, in order to determine potential positions for an operator to stand if they needed to be inside the CBCT room. The following devices were tested: Morita Accuitomo (CBCT1), Newtom Giano HR (CBCT2), Newtom VGi (CBCT3). Scattered radiation measurements were performed using different kVp, mA, and Field of View (FOV) options. An anthropomorphic phantom (NATHANIA) was placed inside the X-ray gantry to simulate clinical conditions. Scattered measurements were taken with the Inovision model 451P Victoreen ionization chamber once placed at fixed distances from each irradiation isocenter, away from the primary beam. A statistically significant (p < 0.001) difference was found in the mean value of the scattered radiation estimations between the CBCT devices. Scattered radiation was reduced with a different rate for each CBCT device as distance was increased. For CBCT1 the reduction was 0.047 μGy, for CBCT2 it was 0.036 μGy, and for CBCT3 it was 0.079 μGy, for every one meter from the X-ray gantry. Therefore, at certain distances from the central X-ray, the scattered radiation was below the critical level of 1 mGy, which is defined by the radiation protection guidelines as the exposure radiation limit of the general population. Consequently, an operator could stay inside the room accompanying the patient being scanned, if necessary.

Publisher

MDPI AG

Subject

General Medicine