Safety of the Radiofrequency Balloon for Pulmonary Vein Isolation: A Focus on Lesion Metric Analysis of Posterior Electrodes

Author:

Almorad Alexandre1,Del Monte Alvise1ORCID,Teumer Yannick2,El Haddad Milad3ORCID,Pannone Luigi1,Della Rocca Domenico Giovanni1ORCID,Audiat Charles1ORCID,Cespón-Fernández María1ORCID,Mouram Sahar1,Ramak Robbert1ORCID,Overeinder Ingrid1,Bala Gezim1,Sorgente Antonio1ORCID,Ströker Erwin1,Sieira Juan1,Brugada Pedro1,La Meir Mark4,de Asmundis Carlo1,Chierchia Gian-Battista1

Affiliation:

1. Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, European Reference Networks Guard-Heart, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, 1090 Brussels, Belgium

2. Department of Medicine II, Ulm University Medical Center, 89070 Ulm, Germany

3. Independent Researcher, 00790 Helsinki, Finland

4. Cardiac Surgery Department, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, 1090 Brussels, Belgium

Abstract

Previous clinical studies on pulmonary vein isolation (PVI) with radiofrequency balloons (RFB) reported safe and effective procedures for a 20 s RF delivery via posterior electrodes. Recent recommendations from the manufacturer suggest reducing the application time to 15 s on the posterior wall (PW) when facing the esophagus region. Here, we retrospectively assess whether 15 s of RF delivery time on posterior electrodes is safe while still ensuring lesion metrics of sufficient quality. This retrospective study included 133 patients with paroxysmal and persistent atrial fibrillation who underwent PVI using an RFB (Heliostar, Biosense Webster, Inc., Irvine, CA, USA) at two European centers. The ablation protocol was set for an RF duration of 20 s/60 s for the posterior/anterior electrodes. A multielectrode temperature probe was systematically used. In the case of an esophageal temperature rise (ETR) above 42 °C (ETR+), an endoscopic evaluation was performed. All posterior electrode lesion metric dynamics (temperature (T) and impedance (Z)) were collected from the RFB generator and analyzed offline. In total, 2435 posterior electrode applications were analyzed. With an RF delivery of 19.8 (19.7–19.8) s, the median impedance drop was 18.4 (12.2–25.2) Ω, while the temperature rise was 11.1 (7.1–14.9) °C. Accordingly, impedance (84.6 (79.3–90.2) Ω) and temperature plateaus (38 (35.3–41.1) °C) were reached at 13.9 (10.6–16) s and 16.4 (12.6–18.5) s, respectively. Overall, 99.6% and 95.8% of electrodes reached 90% (16.6 Ω) and 95% (17.5 Ω) of their impedance drops within 15 s of RF delivery, while 97.2% and 92.8% achieved 90% (34.2 °C) and 95% (36.1 °C) of their temperature rise to reach the plateaus within 15 s of RF delivery. An ETR >42 °C occurred in 37 (30.1%) patients after 17.7 ± 2.3 s of RF delivery. In the ETR+ group, the impedance drop and temperature rise on the posterior electrodes were higher compared to patients where ETR was <42 °C. Two asymptomatic thermal esophageal injuries were observed. In conclusion, 15 s of RF delivery on the posterior electrodes provides a good balance between safety, with no esophageal temperature rise, and efficacy with high-profile lesion metrics.

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3