3D-Printed Biomaterial Testing in Response to Cryoablation: Implications for Surgical Ventricular Tachycardia Ablation

Author:

Candelari Mara1,Cappello Ida Anna1,Pannone Luigi1,Monaco Cinzia1,Bori Edoardo2ORCID,Talevi Giacomo1,Ramak Robbert1ORCID,La Meir Mark3ORCID,Gharaviri Ali1,Chierchia Gian Battista1,de Asmundis Carlo1ORCID,Innocenti Bernardo2ORCID

Affiliation:

1. Heart Rhythm Management Centre, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, European Reference Networks Guard-Heart, 1090 Brussels, Belgium

2. BEAMS Department (Bio Electro and Mechanical Systems), Université Libre de Bruxelles, 1050 Brussels, Belgium

3. Cardiac Surgery Department, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, 1090 Brussels, Belgium

Abstract

Background: The lack of thermally and mechanically performant biomaterials represents the major limit for 3D-printed surgical guides, aimed at facilitating complex surgery and ablations. Methods: Cryosurgery is a treatment for cardiac arrhythmias. It consists of obtaining cryolesions, by freezing the target tissue, resulting in selective and irreversible damage. MED625FLX and TPU95A are two biocompatible materials for surgical guides; however, there are no data on their response to cryoenergy delivery. The study purpose is to evaluate the biomaterials’ thermal properties, examining the temperature changes on the porcine muscle samples (PMS) when the biomaterials are in place during the cryoablation. Two biomaterials were selected, MED625FLX and TPU95A, with two thicknesses (1.0 and 2.5 mm). To analyze the biomaterials’ behavior, the PMS temperatures were measured during cryoablation, firstly without biomaterials (control) and after with the biomaterials in place. To verify the biomaterials’ suitability, the temperatures under the biomaterial samples should not exceed a limit of −30.0 °C. Furthermore, the biomaterials’ geometry after cryoablation was evaluated using the grid paper test. Results: TPU95A (1.0 and 2.5 mm) successfully passed all tests, making this material suitable for cryoablation treatment. MED625FLX of 1.0 mm did not retain its shape, losing its function according to the grid paper test. Further, MED625FLX of 2.5 mm is also suitable for use with a cryoenergy source. Conclusions: TPU95A (1.0 and 2.5 mm) and MED625FLX of 2.5 mm could be used in the design of surgical guides for cryoablation treatment, because of their mechanical, geometrical, and thermal properties. The positive results from the thermal tests on these materials and their thickness prompt further clinical investigation.

Publisher

MDPI AG

Subject

General Medicine

Reference23 articles.

1. Polymeric biomaterials for 3D printing in medicine: An overview;Pugliese;Ann. 3D Print. Med.,2021

2. Cryosurgery for Atrial Fibrillation: Physiologic Basis for Creating Optimal Cryolesions;Cox;Ann. Thorac. Surg.,2021

3. Mechanisms of tissue injury in cryosurgery;Gage;Cryobiology,1998

4. Experimental cryosurgery investigations in vivo;Gage;Cryobiology,2009

5. Cardiac cryosurgery: Effects of myocardial temperature on cryolesion size;Holman;Surgery,1983