Abstract
During the COVID-19 pandemic, fewer in-person clinic visits resulted in fewer point-of-care (POC) HbA1c measurements. In this sub-study, we assessed the performance of alternative glycemic measures that can be obtained remotely, such as HbA1c home kits and Glucose Management Indicator (GMI) values from Dexcom Clarity. Home kit HbA1c (n = 99), GMI, (n = 88), and POC HbA1c (n = 32) were collected from youth with T1D (age 9.7 ± 4.6 years). Bland–Altman analyses and Lin’s concordance correlation coefficient (𝜌c) were used to characterize the agreement between paired HbA1c measures. Both the HbA1c home kit and GMI showed a slight positive bias (mean difference 0.18% and 0.34%, respectively) and strong concordance with POC HbA1c (𝜌c = 0.982 [0.965, 0.991] and 0.823 [0.686, 0.904], respectively). GMI showed a slight positive bias (mean difference 0.28%) and fair concordance (𝜌c = 0.750 [0.658, 0.820]) to the HbA1c home kit. In conclusion, the strong concordance of GMI and home kits to POC A1c measures suggest their utility in telehealth visits assessments. Although these are not candidates for replacement, these measures can facilitate telehealth visits, particularly in the context of other POC HbA1c measurements from an individual.
Funder
National Institute of Diabetes and Digestive and Kidney Diseases
Stanford Diabetes Research Center
Leona M. and Harry B. Helmsley Charitable Trust
Lucile Packard Children’s Hospital Auxiliaries
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献