Abstract
Titanium(IV) solutions are known to detect hydrogen peroxide in solutions by a colorimetric method. Xplosafe’s XploSens PS commercial titanium(IV)-based peroxide detection test strips are used to detect hydrogen peroxide in liquids. The use of these test strips as gas-phase detectors for peroxides was tested using low-cost hardware. The exposure of these strips to hydrogen peroxide liquid or gas leads to the development of an intense yellow color. For liquids, a digital single-lens reflex camera was used to quantify the color change using standardized solutions containing between 50 and 500 ppm peroxide by mass. Analysis of the images with color separation can provide a more quantitative determination than visual comparison to a color chart. For hydrogen peroxide gas, an inexpensive web camera and a tungsten lamp were used to measure the reflected light intensity as a function of exposure from a test strip held in a custom cell. First-order behavior in the color change with time was observed during the exposure to peroxide vapor over a range of peroxide concentrations from 2 and 30 ppm by volume. For a 1-min measurement, the gas-phase detection limit is estimated to be 1 ppm. A 0.01 ppm detection limit can be obtained with a 1-h exposure time. Titanium(IV)-based peroxide detection test strips are sensitive enough to work as a gas-phase hydrogen peroxide detector.