Abstract
Tacrolimus (TAC) is post-transplant pharmacotherapy’s most widely used immunosuppressant. In routine clinical practice, frequent uncomfortable venipuncture is necessary for whole-blood (WB) collection to check trough TAC levels. Volumetric absorptive microsampling (VAMS) is an alternative strategy to WB collection. In this study, we aimed to validate and develop a liquid chromatography–tandem mass spectrometry (LC-MS/MS) method for TAC quantification in WB and VAMS samples. After extraction with water and protein precipitation, the samples were directly analyzed using LC-MS/MS. Whole-blood and VAMS capillary-blood samples were collected from 50 patients treated with TAC during the follow-up visits. The cross-correlation between the developed methods was evaluated using Passing–Bablok regression and a Bland–Altman bias plot. The matrix effect (ME) and carry-over were insignificant for both scenarios. There was a high correlation between the processes and no significant clinical deviation. LC-MS/MS methods were successfully developed and validated in the 0.5–60 ng/mL calibration range. This study demonstrated and confirmed the utility of VAMS-based TAC monitoring in the pediatric population. This is the first study to directly develop and validate the VAMS LC-MS/MS method for evaluating the hematocrit effect in the pediatric population. The statistical correlation between immunochemical and VAMS-based methods was satisfactory.
Funder
Medical University of Warsaw
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献