Abstract
This paper reports a transwell insert-embedded microfluidic device capable of culturing cells at an air-liquid interface (ALI), mimicking the in vivo alveolar epithelium microenvironment. Integration of a commercially available transwell insert makes the device fabrication straightforward and eliminates the tedious device assembly processes. The transwell insert can later be detached from the device for high-resolution imaging of the cells. In the experiments, the cells showing type-I pneumocyte markers are exploited to construct an in vitro alveolar epithelium model, and four culture conditions including conventional liquid/liquid culture (LLC) and air–liquid interface (ALI) cell culture in normal growth medium, and ALI cell culture with inflammatory cytokine (TNF-α) stimulation and ethanol vapor exposure are applied to investigate their effects on the alveolar epithelium barrier function. The barrier permeability is time-lapse monitored using trans-epithelial electrical resistance (TEER) measurement and immunofluorescence staining of the tight junction protein (ZO-1). The results demonstrate the functionalities of the device, and further show the applications and advantages of the constructed in vitro cell models for the lung studies.
Funder
Ministry of Science and Technology, Taiwan
Academia Sinica
Subject
Electrical and Electronic Engineering,Mechanical Engineering,Control and Systems Engineering
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献