Assessment of Shoulder Range of Motion Using a Wireless Inertial Motion Capture Device—A Validation Study

Author:

Rigoni ,Gill ,Babazadeh ,Elsewaisy ,Gillies ,Nguyen ,Pathirana ,Page

Abstract

(1) Background: Measuring joint range of motion has traditionally occurred with a universal goniometer or expensive laboratory based kinematic analysis systems. Technological advances in wearable inertial measurement units (IMU) enables limb motion to be measured with a small portable electronic device. This paper aims to validate an IMU, the ‘Biokin’, for measuring shoulder range of motion in healthy adults; (2) Methods: Thirty participants completed four shoulder movements (forward flexion, abduction, and internal and external rotation) on each shoulder. Each movement was assessed with a goniometer and the IMU by two testers independently. The extent of agreement between each tester’s goniometer and IMU measurements was assessed with intra-class correlation coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA). Secondary analysis compared agreement between tester’s goniometer or IMU measurements (inter-rater reliability) using ICC’s and LOA; (3) Results: Goniometer and IMU measurements for all movements showed high levels of agreement when taken by the same tester; ICCs > 0.90 and LOAs < ±5 degrees. Inter-rater reliability was lower; ICCs ranged between 0.71 to 0.89 and LOAs were outside a prior defined acceptable LOAs (i.e., > ±5 degrees); (4) Conclusions: The current study provides preliminary evidence of the concurrent validity of the Biokin IMU for assessing shoulder movements, but only when a single tester took measurements. Further testing of the Biokin’s psychometric properties is required before it can be confidently used in routine clinical practice and research settings.

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Reference28 articles.

1. The reliability of clinical measurements designed to quantify shoulder mobility

2. Validity of Goniometric Elbow Measurements: Comparative Study with a Radiographic Method

3. The reliability and concurrent validity of shoulder mobility measurements using a digital inclinometer and goniometer: A technical report;Kolber;Int. J. Sports Phys. Ther.,2012

4. Upper limb motion tracking with the integration of IMU and Kinect

5. Measurement of Human Daily Physical Activity

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3