Abstract
A new electrochemical sensor designed by modifying the commercial boron-doped diamond (BDD) electrode with graphene oxide (GO) reduced electrochemically and further electrodecorated with silver (Ag), named the Ag/GO/BDD electrode, was selected among a series of the BDD, GOelectroreduced onto BDD (GO/BDD) and silver electrodeposited onto BDD (Ag/BDD) electrodes for the detection of tetracycline (TC) in aqueous solution. The best results regarding the sensitivity of 46.6 µA·µM−1·cm−2 and the lowest limit of detection of 5 nM was achieved using square-wave voltammetry (SWV) operated at the step potential of 5 mV, modulation amplitude of 200 mV and the frequency of 10 Hz in alkaline medium. The application of the alkaline supporting electrolyte-based procedure is limited for water monitoring due to the presence of chloride that interferes with TC detection; however, it can be applied for quantitative determination of pharmaceutical formulations. 0.1 M Na2SO4 supporting electrolyte eliminated chloride interference and can be used for the application of Ag/GO/BDD in practical detection of TC in water.
Subject
General Materials Science,General Chemical Engineering
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献