Abstract
This study compared hematocrit measured with the EPOC and i-STAT Alinity v point-of-care analyzers and manual measurement of packed cell volume in managed African savanna elephants (Loxodonta africana) and southern white rhinoceros (Ceratotherium simum simum). Biochemical analytes were also measured with the EPOC, i-STAT Alinity v, and a veterinary diagnostic laboratory in the same animals. Analytes assessed included blood urea nitrogen, chloride, creatinine, glucose, ionized calcium, potassium, and sodium. There were no differences for hematocrit values for African savanna elephants or southern white rhinoceros (p ≤ 0.05). In African savanna elephants, there were no differences between the EPOC and i-STAT Alinity v analyzers for any measured analytes except ionized calcium. When compared to a veterinary diagnostic laboratory, there were differences for a majority of the biochemical analytes measured on the EPOC and i-STAT Alinity v analyzers in African savanna elephants. In southern white rhinoceros, there were differences for a majority of analytes among all three analyzers. While differences existed among the portable analyzers and a veterinary diagnostic laboratory for biochemical analytes in both species, these numerically small differences are unlikely to be clinically significant. For routine health care of African savanna elephants and southern white rhinoceros, these point-of-care analyzers may be a useful alternative to commercial analyzers for the parameters evaluated.