Intradevice Repeatability and Interdevice Comparison of Two Specular Microscopy Devices in a Real-Life Setting: Tomey EM-4000 and Nidek CEM-530

Author:

Kecik Mateusz12,Kropp Martina12ORCID,Thumann Gabriele12,Pajic Bojan12345ORCID,Guber Josef6,Guber Ivo1ORCID

Affiliation:

1. Department of Ophthalmology, University of Geneva, 1205 Geneva, Switzerland

2. Experimental Ophthalmology, University of Geneva, 1205 Geneva, Switzerland

3. Eye Clinic ORASIS, Swiss Eye Research Foundation, 5734 Reinach, Switzerland

4. Faculty of Sciences, Department of Physics, University of Novi Sad, Trg Dositeja Obradovica 4, 21000 Novi Sad, Serbia

5. Faculty of Medicine of the Military Medical Academy, University of Defense, 11000 Belgrade, Serbia

6. Department of Ophthalmology, University of Basel, 4001 Basel, Switzerland

Abstract

Background and Objectives: The purpose of this study was to compare two commercially available specular microscopes (Tomey EM-4000 and Nidek CEM-530) in a real-life clinical setting in terms of intra- and interdevice variability. The study was conducted on all patients seen in a clinical practice specializing in anterior segment pathologies, regardless of the purpose of their visit. Materials and Methods: In total, 112 eyes of 56 patients (age 23–85 years old) were included in the study. Each eye was measured three times with each device (for a total of six measurements), and results for central corneal thickness (CCT) and corneal endothelial cell density (ECD) were recorded. The results were then evaluated with the D’Agostino–Pearson normality test and compared with a Wilcoxon signed-rank test, t-test, ANOVA or Mann–Whitney test for intra- and interdevice variability. Results: Both specular microscopes produced very reliable reproducible intradevice results: The Tomey EM-4000 measured an ECD of 2390 ± 49.57 cells/mm2 (mean ± standard error of mean); the range was 799–3010 cells/mm2. The determined CCT was 546 ± 5.104 µm (mean ± standard error of mean [SEM]); the range was 425–615 µm. The measurements with the Nidek CEM-530 revealed an ECD of 2417 ± 0.09 cells/mm2 (mean ± SEM); the range was 505–3461 cells/mm2 (mean ± SEM). The mean CCT detected was 546.3 ± 4.937 µm (mean ± SEM); the range was 431–621 µm. The interdevice differences were statistically significant for both parameters, ECD (p = 0.0175) and CCT (p = 0.0125) (p < 0.05). Conclusions: The Nidek CEM-530 and the Tomey EM-4000 both produced reliable and reproducible results in terms of ECD and CCT. The absolute measurements were statistically significantly different for CCT and ECD for both devices; the Nidek produces slightly higher values.

Publisher

MDPI AG

Reference34 articles.

1. Molecular mechanisms underlying the corneal endothelial pump;Bonanno;Exp. Eye Res.,2012

2. Overview of the Cornea: Structure, Function, and Development;Eghrari;Prog. Mol. Biol. Transl. Sci.,2015

3. The balance between corneal transparency and edema: The Proctor Lecture;Edelhauser;Investig. Ophthalmol. Vis. Sci.,2006

4. The resiliency of the corneal endothelium to refractive and intraocular surgery;Edelhauser;Cornea,2000

5. Changes in the normal corneal endothelial cellular pattern as a function of age;Yee;Curr. Eye Res.,1985

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3