A Novel Technique for Identification of Wear Values at Different Lengths after Multiple Clinical Use of Different File Systems

Author:

Çırakoğlu Neslihan YılmazORCID,Çiçek Ersan,Özarpa Cevat,Özbay Yağız,Özdemir OlcayORCID

Abstract

Introduction The purpose of this research is to assess the wearing of the rotary file system (Protaper Next) and reciprocating file systems (Reciproc Blue and WaveOne Gold) at different lengths using a novel technique after in vivo clinical use. Materials and Methods Twelve different unused samples from each brand were accepted as reference values. For three different brands, the diameters of the files were measured by taking 12 samples used once, 12 samples used twice, and 12 samples used three times. Images were taken with a USB Micron Microscope, and file diameters were measured by determining limit values with Autocad. Result Reciproc Blue system was the most worn at apical 1 mm, and WaveOne Gold system was the most worn at apical 3 mm. PTN system exhibited the least wearing at any length. Moreover, less wearing was observed in the rotation motion than in the reciprocating motion. Conclusion In clinical practice, for the guttapercha to be fully adapted to the apical construction prepared according to the determined WL, the file should not undergo any wearing and volume reduction. Wearing—especially in the apical parts of the file—causes less preparation, and this situation could lead to apically obturation failure.

Funder

Karabük University Scientific Research Projects Office, TURKEY

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3