Comparison of Preoperative ECG Screening and Device-Based Vector Analysis in Patients Receiving a Subcutaneous Implantable Cardioverter-Defibrillator

Author:

Budrejko Szymon1ORCID,Zienciuk-Krajka Agnieszka1ORCID,Daniłowicz-Szymanowicz Ludmiła1ORCID,Kempa Maciej1

Affiliation:

1. Department of Cardiology and Electrotherapy, Faculty of Medicine, Medical University of Gdansk, Smoluchowskiego 17, 80-214 Gdansk, Poland

Abstract

Background and Objectives: Subcutaneous implantable cardioverter-defibrillators (S-ICDs) provide protection against sudden cardiac death from outside the cardiovascular system. ECG screening is a prerequisite for implantation, but the reproducibility of its results post-operatively in the device is only partial. We aimed to compare the results of ECG screening with device-based sensing vector analysis. Materials and Methods: We screened the hospital records of all S-ICD recipients in our clinic. All of them had pre-operative ECG screening performed (primary, secondary, and alternate vectors). The results were compared with device-based vector analysis to determine the relation of the pre- and post-operative vector availability. Results: Complete ECG screening and device-based vector analysis were obtained for 103 patients. At least two acceptable vectors were found in 97.1% of the patients pre-operatively and in 96.1% post-operatively. When comparing vectors in terms of agreement (OK or FAIL) pre- and post-operatively, in 89.3% of the patients, the result for the primary vector was the same in both situations; for the secondary, it was in 84.5%, and for the alternate, it was in 74.8% of patients, respectively. In 55.3% of patients, all three vectors were labeled the same (OK or FAIL); in 37.9%, two vectors had the same result, and in 6.8%, only one vector had the same result pre- and post-operatively. The number of available vectors was the same pre- and post-operatively in 62.1% of patients, while in 15.5%, it was lower, and in 22.3% of patients, it was higher than observed during screening. Conclusions: Routine clinical pre-operative screening allowed for a good selection of candidates for S-ICD implantation. All patients had at least one vector available post-operatively. The final number of vectors available in the device-based analysis in most patients was at least the same (or higher) than during screening. The repeatability of the positive result for a single vector was high.

Publisher

MDPI AG

Subject

General Medicine

Reference52 articles.

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3