Accuracy and Reliability of Software Navigation for Acetabular Component Placement in THA: An In Vitro Validation Study

Author:

Brady Alex W.,Tatka Jakub,Fagotti Lorenzo,Kemler Bryson R.,Fossum Bradley W.ORCID

Abstract

Background and Objectives: Intraoperative fluoroscopy can be used to increase the accuracy of the acetabular component positioning during total hip arthroplasty. However, given the three-dimensional nature of cup positioning, it can be difficult to accurately assess inclination and anteversion angles based on two-dimensional imaging. The purpose of this study is to validate a novel method for calculating the 3D orientation of the acetabular cup from 2D fluoroscopic imaging. Materials and Methods: An acetabular cup was implanted into a radio-opaque pelvis model in nine positions sequentially, and the inclination and anteversion angles were collected in each position using two methods: (1) a coordinate measurement machine (CMM) was used to establish a digitalized anatomical coordinate frame based on pelvic landmarks of the cadaveric specimen, and the 3D position of the cup was then expressed with respect to the anatomical planes; (2) AP radiographic images were collected, and a mathematical formula was utilized to calculate the 3D inclination and anteversion based on the 2D images. The results of each method were compared, and interrater and intrarater reliably of the 2D method were calculated. Results: Interrater reliability was excellent, with an interclass correlation coefficient (ICC) of 0.988 (95% CI 0.975–0.994) for anteversion and 0.997 (95% CI 0.991–0.999) for inclination, as was intrarater reliability, with an ICC of 0.995 (95% CI 0.985–0.998) for anteversion and 0.998 (95% CI 0.994–0.999) for inclination. Intermethod accuracy was excellent with an ICC of 0.986 (95% CI: 0.972–0.993) for anteversion and 0.993 (95% CI: 0.989–0.995) for inclination. The Bland–Altman limit of agreement, which represents the error between the 2D and 3D methods, was found to range between 2 to 5 degrees. Conclusions: This data validates the proposed methodology to calculate 3D anteversion and inclination angles based on 2D fluoroscopic images to within five degrees. This method can be utilized to improve acetabular component placing intraoperatively and to check component placement postoperatively.

Funder

DePuy Synthes

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3