Accuracy Validation of the Elecsys HBsAg II Quant Assay and Its Utility in Resolving Equivocal Qualitative HBsAg Results

Author:

Lee Jaehyeon123ORCID,Lee Seung Yeob123,Cho Yong Gon123,Kim Dal Sik123,Park Joonhong123ORCID

Affiliation:

1. Department of Laboratory Medicine, Jeonbuk National University Medical School and Hospital, Jeonju 54907, Republic of Korea

2. Research Institute of Clinical Medicine, Jeonbuk National University, Jeonju 54907, Republic of Korea

3. Biomedical Research Institute, Jeonbuk National University Hospital, Jeonju 54907, Republic of Korea

Abstract

Background and Objectives: There are reports of false qualitative HBsAg results, because of various causes, such as samples with low HBsAg concentrations that may produce false positives. The main aims of this study were to validate the analytical accuracy and to assess the utility of the Elecsys assay compared to that of the qualitative HbsAg assay as a screening test in resolving equivocal qualitative HbsAg results. Materials and Methods: The limit of blank (LoB), the limit of detection (LoD), the limit of quantification (LoQ), and linearity were estimated to validate the analytical accuracy of the Elecsys HBsAg II Quant assay. A total of 449 serum samples showing initial equivocal results (1–50 index) were evaluated by Elecsys HBsAg II Quant and ADVIA Centaur HBsAg II assays. Results: The LoQ of the assay was determined to be 0.050 IU/mL, as provided by the manufacturer. The Kappa agreement between the two assays was almost perfect, at 0.9669, despite seven discordant results. With a specificity of 100% at new cut-off index value ≥5.42, about 78 samples (17%, 78/449) with index value ≥5.42 were interpreted as positives without further duplicate tests, however the remaining 371 samples with index value <5.42 need to be confirmed with additional HBV marker assays. Conclusions: We confirm that the Elecsys HBsAg II Quant assay is accurate and sensitive for HBV infection and recommend it as an alternative confirmatory HBsAg assay for resolving equivocal qualitative HBsAg results.

Funder

Roche Diagnostics Korea Co., Ltd., Seoul, Republic of Korea

Publisher

MDPI AG

Subject

General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3