Topical Anaesthesia Using a Soft Mist Spray Device Allows Comfortable Awake Visualisation of the Airway via Self-Videolaryngoscopy in Volunteers

Author:

Markerink Hielke1ORCID,van Geffen Geert-Jan1,Bruhn Jörgen1

Affiliation:

1. Department of Anesthesiology, Pain and Palliative Medicine, Radboud University Medical Centre (RUMC), 6525 GA Nijmegen, The Netherlands

Abstract

Background: During endotracheal intubation, there is a 10% incidence of difficult laryngoscopy, which may result in serious complications. It is important to obtain as much information about the visibility of laryngeal structures before the patient is anaesthetised. Performing awake (video-) laryngoscopy on a patient is uncomfortable and can trigger gagging and coughing reflexes, making visualisation nearly impossible. The objective of this study is to evaluate the effectiveness of a soft mist spray device for airway anaesthesia during awake (video-) laryngoscopy. Methods: Twenty healthy volunteers inhaled through the Trachospray device, which was placed in their mouths. Two 2 mL syringes containing lidocaine at 4% were sprayed into the airway during inspiration. After several minutes, the subjects were asked to perform a videolaryngoscopy on themselves until the glottic structures and the vocal cords were visible. Upon completion of the procedure, all participants were asked to fill out a feedback form. Results: The duration of the videolaryngoscopy to visualisation of the vocal cords averaged 17 ± 13 s. After analysing the data, three distinct groups emerged as follows: Group 1 (70% of participants) showed no response, allowing for easy insertion of the videolaryngoscope. Group 2 (25% of participants) exhibited a light response but still permitted easy insertion and visualisation. One patient demonstrated a clear response with noticeable laryngeal contraction, requiring slightly more effort and discomfort for insertion. In 80% of the participants, the laryngeal structures were visualised according to Cormack–Lehane grade 1. All participants reported a high level of comfort, with an average rating of NRS 8. The anaesthesiologist assessed the level of anaesthesia as good to very good. No adverse events were observed. Conclusions: The Trachospray provided good, reliable, comfortable, and safe topical anaesthesia for awake videolaryngoscopy. This enables a direct visual assessment of the airway and may assist in making decisions regarding airway management for tracheal intubation.

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3