Abstract
Objective: The aim of this study was to evaluate the influence of autoclave sterilization on the resistance to cyclic fatigue of two nickel-titanium (NiTi) endodontic files of identical design and taper, but with different NiTi alloy treatments: the newly introduced heat-treated Race Evo and the electropolished Race files. Materials and methods: Fifteen Race (25/0.06) files and fifteen Race Evo (25/0.06) files (n = 30 in total) were randomly assigned to five sub-groups each consisting of three files of the same NiTi alloy treatment. One group served as a control with files unautoclaved. The four remaining groups were sterilized in a steam sterilizer for 1, 3, 5, and 10 autoclave cycles, respectively. Files then underwent cyclic fatigue testing in a simulated metal canal block. A scanning electron microscope was used to inspect the surface of the fractured instruments. Statistical analysis was conducted using independent t-test and multi-factorial analysis of variance with significance set at a p value of ≤0.05. Results: Both Race Evo and Race files showed no significant difference between the different autoclaving cycles in terms of the number of cycles to fracture (p = 0.232 and p = 0.359). Despite rotating at a higher speed, the number of cycles to fracture of heat-treated Race Evo files was significantly higher than that of Race files (p ≤ 0.0001). Conclusion: Autoclave sterilization has no significant effect on the resistance to cyclic fatigue of heat-treated Race Evo or electropolished Race files. However, Race Evo files showed superior resistance to cyclic fatigue irrespective of autoclaving cycles.
Subject
General Materials Science,Metals and Alloys
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献