Hydrophilic Modification of Dialysis Membranes Sustains Middle Molecule Removal and Filtration Characteristics

Author:

Zawada Adam M.1ORCID,Emal Karlee2ORCID,Förster Eva1,Saremi Saeedeh13ORCID,Delinski Dirk1,Theis Lukas1,Küng Florian1,Xie Wenhao4,Werner Joanie5,Stauss-Grabo Manuela6,Faust Matthias3,Boyington Skyler2,Kennedy James P.7

Affiliation:

1. Product Development, Fresenius Medical Care Deutschland GmbH, 66606 Sankt Wendel, Germany

2. Applications Laboratory, Fresenius Medical Care, Ogden, UT 84404, USA

3. Institute for Physical Process Technology, Saarland University of Applied Sciences, 66117 Saarbrücken, Germany

4. Product Development, Fresenius Medical Care, Shanghai 200233, China

5. Clinical Marketing & Innovations, Fresenius Medical Care, Waltham, MA 02451, USA

6. Global Biomedical Evidence Generation, Fresenius Medical Care Deutschland GmbH, 61352 Bad Homburg, Germany

7. Product Development, Fresenius Medical Care, Ogden, UT 84404, USA

Abstract

While efficient removal of uremic toxins and accumulated water is pivotal for the well-being of dialysis patients, protein adsorption to the dialyzer membrane reduces the performance of a dialyzer. Hydrophilic membrane modification with polyvinylpyrrolidone (PVP) has been shown to reduce protein adsorption and to stabilize membrane permeability. In this study we compared middle molecule clearance and filtration performance of nine polysulfone-, polyethersulfone-, and cellulose-based dialyzers over time. Protein adsorption was simulated in recirculation experiments, while β2-microglobulin clearance as well as transmembrane pressure (TMP) and filtrate flow were determined over time. The results of this study showed that β2-microglobulin clearance (−7.2 mL/min/m2) and filtrate flow (−54.4 mL/min) decreased strongly during the first 30 min and slowly afterwards (−0.7 mL/min/m2 and −6.8 mL/min, respectively, for the next 30 min); the TMP increase (+37.2 mmHg and +8.6 mmHg, respectively) showed comparable kinetics. Across all tested dialyzers, the dialyzer with a hydrophilic modified membrane (FX CorAL) had the highest β2-microglobulin clearance after protein fouling and the most stable filtration characteristics. In conclusion, hydrophilic membrane modification with PVP stabilizes the removal capacity of middle molecules and filtration performance over time. Such dialyzers may have benefits during hemodiafiltration treatments which aim to achieve high exchange volumes.

Publisher

MDPI AG

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3