Abstract
An area detector with a central hole structure was built up for small-angle neutron scattering (SANS) on the iMATERIA instrument at Japan Proton Accelerator Research Complex (J-PARC). Linear position-sensitive detector tubes filled with 3He gas were arranged in three layers leaving a central hole. As a result of the calibration process, a SANS measurement with wide q-range from 0.007 Å−1 to 4.3 Å−1 was achieved in double-frame operation, supplying neutrons with wavelengths from 1 Å to 10 Å. As a merit of this central hole structure, neutron transmission can be measured simultaneously to reduce experimental time and effort. This is ideal for time-resolved studies, in which the sample transmission can be time-dependent, throughout the whole experiment. Additionally, the data storage system in ‘event mode’ format provides an excellent platform for such time-resolved experiments.