Self-Isolated Dual-Mode High-Pass Birdcage RF Coil for Proton and Sodium MR Imaging at 7 T MRI

Author:

Jeong You-Jin1ORCID,Kumar Suchit2ORCID,Park Seon-Woo2,Song Youngkyu3,Cho Jee-Hyun3ORCID,Moon Chan Hong4,Oh Chang-Hyun12

Affiliation:

1. Department of Electronics and Information Engineering, Korea University, Sejong 30019, Republic of Korea

2. Research Institute for Advanced Industrial Technology, Korea University, Sejong 30019, Republic of Korea

3. Bio-Chemical Analysis Team, Korea Basic Science Institute, Cheongju 28119, Republic of Korea

4. Department of Radiology, University of Pittsburgh, Pittsburgh, PA 15213, USA

Abstract

This study presents the feasibility of a dual-mode high-pass birdcage RF coil to acquire MR images at both 1H and 23Na frequencies at ultra-high-field MR scanner, 7 T. A dual-mode circuit (DMC) in the dual-mode birdcage (DMBC) RF coil operates at two frequencies, addressing the limitations of sensitivity reduction and isolation between two frequencies as in traditional dual-tuned RF coil. Finite-difference time-domain (FDTD) based electromagnetic (EM) simulations were performed to verify the RF coil at each frequency on the three-dimensional human head model. The DMBC RF coil resonated at proton (1H) and sodium (23Na) frequencies, and also single-tuned high-pass birdcage RF coils were constructed for both 1H and 23Na frequencies. The bench test performance of the RF coils was evaluated using network analysis parameters, including the measurement of scattering parameters (S-parameters) and quality factors (Q-factors). Q-factor of the DMBC coil at 1H port was 10.2% lower than that of 1H single-tuned birdcage (STBC) coil, with a modest SNR reduction of 6.5%. Similarly, the Q-factor for the DMBC coil at 23Na port was 12.3% less than that of 23Na STBC coil, and the SNR showed a minimal reduction of 5.4%. Utilizing the DMBC coil, promising 1H and 23Na MR images were acquired compared to those by using STBC coils. In conclusion, deploying a DMBC 1H/23Na coil has been demonstrated to overcome traditional constraints associated with dual-tuned RF coils, achieving this with only nominal signal attenuation across both nuclei operational frequencies.

Funder

Institute of Information & Communications Technology Planning & Evaluation

Publisher

MDPI AG

Subject

Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A Modified Quadrature Birdcage Coil Incorporated With a Curved Feature for In Ovo MR Imaging;IEEE Open Journal of Engineering in Medicine and Biology;2024

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3