Abstract
Absolute measurement of radiant power in the X-ray region is essential for many applications in astrophysics, spectroscopy, and X-ray diagnostics. Comparison between different measuring methods is an effective way to check their reliability. In the present work, a comparison of X-ray radiant power absolute measurement between a free-air ionization chamber and a cryogenic electrical substitution radiometer was performed at Beijing Synchrotron Radiation Facility. The absolute radiant power obtained by these two methods were mutually compared via a transfer standard detector’s spectral responsivity at a photon energy of 10 keV. The result of the comparison showed that the difference was 0.47%. A conclusion was reached that the free-air ionization chamber and the cryogenic electrical substitution radiometer agreed within the combined relative uncertainty of 3.35%.
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry