Abstract
This paper presents a power line interference (PLI) reduction technique with a current-reused current-feedback instrumentation amplifier (CFIA) for electrocardiogram (ECG) recording. In a portable two-electrode ECG monitoring application, the presence of undesired PLI may severely corrupt the quality of ECG recording. Since PLI can be over a few volts, the input signal including the ECG signal can exceed the supply or ground level by an electrostatic discharge (ESD) diode in input/output (I/O) pad. To prevent this problem, this paper presents a continuous-time input common-mode current feedback loop that can limit displacement current from a capacitive coupling between the human body and a power line. The continuous-time input common-mode current feedback loop can clamp an input common-mode voltage to the saturation region of the input transistor of the current-reused CFIA. After the clamping procedure, the clamped input signal is amplified by the current-reused CFIA. The proposed circuit was designed using a 0.18-μm bipolar-complementary metal semiconductor–double-diffused metal oxide semiconductor (BCDMOS) process with an active area of 1.8 mm2. The total power consumption is 18 μW with 1.8 V. The input-referred noise and noise efficiency factor (NEF) of the current-reused CFIA is 2.68 μVRMS and 4.28 with 107 Hz, respectively.
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献