Abstract
In this research, a boron-doped diamond (BDD) electrode has been explored to detect the chloroquine drug. The electrochemical performance of BDD electrode towards the irreversible anodic response of chloroquine was investigated by subjecting this electrode to the cathodic (−0.5 A cm−2 by 180 s, generating a predominantly hydrogen-terminated surface) and anodic (+0.5 A cm−2 by 30 s, oxygen-terminated surface) pretreatments. The cathodically pretreated BDD electrode ensured a better-defined anodic peak and higher current intensity. Thus, by applying the cathodically pretreated BDD electrode and square-wave voltammetry (SWV), the analytical curve was linear from 0.01 to 0.25 µmol L−1 (correlation coefficient of 0.994), with sensitivity and limit of detection of 12.2 µA L µmol−1 and 2.0 nmol−1, respectively. This nanomolar limit of detection is the lowest recorded so far with modified and unmodified electrodes.
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献