Abstract
The design and synthesis of artificial receptors based on molecular imprinting (MI) technology for the development of a new MIP-based biosensor for detection of the stress biomarker α-amylase in human saliva in point-of-care (PoC) applications is described in this work. The portable electrochemical devices for monitoring α-amylase consists of cost-effective and disposable gold screen-printed electrodes (AuSPEs). To build the electrochemical device, the template biomolecule was firstly immobilized directly over the working area of the gold chip previously activated with a self-assembled monolayer (SAM) of cysteamine (CA). Then, pyrrole (Py) monomer was selected as building block of a polymeric network prepared by CV electropolymerization. After the electropolymerization process, the enzyme was removed from the polymer film in order to build the specific recognition sites for the target enzyme. The MIP biosensor showed a very wide linear concentration range (between 3.0 × 10−4 to 0.60 mg mL−1 in buffer solution and between 3.0 × 10−4 to 3.0 × 10−2 mg mL−1 in human saliva) and low detection levels were achieved (LOD < 3.0 × 10−4 mg mL−1) using square wave voltammetry (SWV) as the electroanalytical technique.
Funder
Fundação para a Ciência e a Tecnologia
Cited by
19 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献