Abstract
This study aimed to compare the intubation effectiveness of the bébé Vie Scope™ (VieScope) and direct laryngoscopy for emergency intubation in a pediatric manikin model performed by paramedics with and without personal protective equipment for aerosol generating procedures (PPE-AGP). Participants performed endotracheal intubation using VieScope and standard Macintosh laryngoscope (MAC) in two research scenarios: (1) without PPE-AGP, and (2) with PPE-AGP. Fifty-one paramedics without any previous experience with the VieScope participated in this study. In the PPE-AGP scenario, in the VieScope group, the percentage of successful tracheal intubation on the first attempt was higher compared to the MAC group (94.1 vs. 78.4%, p = 0.031), intubation time was shorter (29.8 vs. 33.9 s, p < 0.001), and percentage of glottic opening (POGO) score was higher 91.0 vs 77.8 (p < 0.001). On the Cormack–Lehane scale, intubation with VieScope intubation was associated with higher scores rated at 1 (64.7 vs. 29.4%) than in the MAC group (p = 0.001). For intubation in the non-PPE scenario, there were no statistically significant differences between VieScope and MAC in relation to above parameters. Summarize, the bébé VieScope™ under PPE-AGP wearing conditions has proven to be a useful device for airway management in children providing better visualization of the larynx, better intubation conditions, and a higher success rate of tracheal intubation on the first attempt and reduced intubation time compared to the standard Macintosh laryngoscope.