Does Focal Osteolysis in a PRECICE Stryde Intramedullary Lengthening Nail Resolve after Explantation?

Author:

Sax Oliver C.,Hlukha Larysa P.,Kowalewski Kyle A.,Herzenberg John E.,McClure Philip K.

Abstract

Concerns surrounding osteolysis near and around the modular junction of a stainless-steel intramedullary lengthening rod prompted a manufacturer recall from the United States market in early 2021. These actions were preceded by similar steps taken in Europe. A concomitant review of stainless-steel lengthenings at our institution demonstrated signs of adverse tissue reaction including periosteal reaction and osteolysis at the modular junction and/or male-sided locking screws. Nearly half of our patients presented with these findings on radiographic images. At the time of the previous review, only half of the nearly 60 implanted stainless-steel devices met a 6-month follow-up. At this juncture, many patients have had their devices explanted. Given the suspected adverse tissue reactions caused by a component of the internal device, we sought to examine the rate of osteolysis post-explantation following removal of a stainless-steel nail. We reviewed a consecutive series of patients who underwent implantation of a stainless-steel limb lengthening device in the femur and/or tibia at a single institution between December 2018 and December 2020. Patients were included if their device was explanted. Periosteal reaction and osteolysis was classified according to a novel and validated classification system, as analyzed by five fellowship-trained surgeons. In addition, changes observed prior to explantation were tracked post-explantation to assess for resolution. The incidence of periosteal reaction and osteolysis prior to explantation was 22/57 (39%) and 15/57 (26%), respectively. Of the 15 patients with osteolysis pre-explantation, 14 patients’ implants were explanted. Of these, eight patients had available follow-up films. Two patients were identified as having partial osteolysis resolution at mean 1-year follow-up, while six patients were identified as having complete osteolysis at mean 18-months follow-up. Periosteal tissue reaction and osteolysis largely resolved following explantation in a subset of patients. These results provide further support to the claim that the stainless-steel device contributed to the changes seen. Further follow-up is warranted to examine the longer-term effects of adverse tissue reaction in this patient population.

Publisher

MDPI AG

Subject

Pediatrics, Perinatology and Child Health

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3