A 2-month-old male 1.56 kg Yorkshire terrier (Case No. 1) and a 3-month-old male 2.3 kg Jack Russell Terrier (Case No. 2) were scheduled for ophthalmological surgery under general anaesthesia and neuromuscular blockade. For both patients, volume-controlled ventilation (VCV) was used with set tidal volumes (VT) of 13 mL/kg and 20 mL/kg for cases No. 1 and 2, respectively. The type of ventilator used did not take into account the intrinsic compliance of the breathing system; therefore, a significant part of the delivered VT was wasted in the expansion of the breathing system, and did not reach the patients, causing alveolar hypoventilation. Both cases developed low dynamic compliance (CD), and after a recruitment manoeuvre, EtCO2 of up to 116 mmHg and 197 mmHg were revealed for cases No. 1 and 2, respectively. The two cases had to be ventilated manually, using positive inspiratory pressures (PIP) of 20–25 mmHg, in order to improve alveolar ventilation and reduce the EtCO2, as adjustments to the VCV were ineffective. Both patients maintained an oxygen haemoglobin saturation between 94% and 100% throughout the procedure and they recovered well. Using a higher VT from the beginning, to compensate for the compliance of the breathing system, or the use of pressure-controlled ventilation (PCV), could have potentially helped to avoid these two incidences of severe hypercapnia.