Abstract
Transradial coronaro-angiography (TRA) can be performed with one catheter. We investigate the efficacy of four different DxTerity catheter curves dedicated to the single-catheter technique and compare this method to the standard two-catheter approach. For this prospective, single-blinded, randomized pilot study, we enrolled 100 patients. In groups 1, 2, 3, and 4, the DxTerity catheters Trapease, Ultra, Transformer and Tracker Curve, respectively, were used. In group 5 (control), standard Judkins catheters were used. The study endpoints were the percentage of optimal stability, proper ostial artery engagement and a good quality angiogram, the duration of each procedure stage, the amount of contrast, and the radiation dose. The highest rate of optimal stability was observed in groups 2 (90%) and 5 (95%). Suboptimal results with at least one episode of catheter fallout from the ostium were most frequent in group 1 (45%). The necessity of using another catheter was observed most frequently in group 4. The analysis of time frames directly depending on the catheter type revealed that the shortest time for catheter introduction and for searching coronary ostia was achieved in group 2 (Ultra). There were no differences in contrast volume and radiation dose between groups. DxTerity catheters are suitable tools to perform TRA coronary angiography. The Ultra Curve catheter demonstrated an advantage over other catheters in terms of its ostial stability rate and procedural time.
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献