Author:
Gil Bruno,Anastasova Salzitsa,Yang Guang
Abstract
Wearable biomedical technology has gained much support lately as devices have become more affordable to the general public and they can easily interact with mobile phones and other platforms. The feasibility and accuracy of the data generated by these devices so as to replace the standard medical methods in use today is still under scrutiny. In this paper, we present an ear-worn device to measure cardiovascular and sweat parameters during physical exercise. ECG bipolar recordings capture the electric potential around both ears, whereas sweat rate is estimated by the impedance method over one segment of tissue closer to the left ear, complemented by the measurement of the lactate and pH levels using amperiometric and potentiometric sensors, respectively. Together with head acceleration, the acquired data is sent to a mobile phone via BLE, enabling extended periods of signal recording. Results obtained by the device have shown a SNR level of 18 dB for the ECG signal recorded around the ears, a THD value of −20.46 dB for the excitation signal involved in impedance measurements, sweat conductivity of 0.08 S/m at 1 kHz and sensitivities of 50 mV/pH and 0.8 μA/mM for the pH and lactate acquisition channels, respectively. Testing of the device was performed in human subjects during indoors cycling with characteristic level changes.
Funder
Engineering and Physical Sciences Research Council
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
48 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献