Abstract
For years, capacitive electrocardiogram (CECG) has been known to be susceptible to ambient interference. In light of this, a novel capacitive electrode was developed as an effective way to reduce the interference effect. This was done by simply introducing the capacitive elector in series with a 1 pF capacitor, and the 60 Hz common mode noise induced by AC power lines was cancelled using a capacitive right leg (CRL) circuit. The proposed electrode did as expected outperform two counterparts in terms of SNR, and particularly gave an up to 99.8% correlation between RRIs extracted from an ECG and a CECG signal, a figure far beyond 52% and 63% using the two counterparts. This capacitive electrode was originally designed for long-term noncontact monitoring of heart rate, and hopefully can be integrated to portable devices for other medical care services in the near future.
Funder
Ministry of Science and Technology, Taiwan
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
3 articles.