Abstract
X-ray free electron lasers, with their ultrashort highly coherent pulses, opened up the opportunity of probing ultrafast nano- and atomic-scale dynamics in amorphous and disordered material systems via speckle visibility spectroscopy. However, the anticipated count rate in a typical experiment is usually low. Therefore, visibility needs to be extracted via photon statistics analysis, i.e., by estimating the probabilities of multiple photons per pixel events using pixelated detectors. Considering the realistic X-ray detector responses including charge cloud sharing between pixels, pixel readout noise, and gain non-uniformity, speckle visibility extraction relying on photon assignment algorithms are often computationally demanding and suffer from systematic errors. In this paper, we present a systematic study of the commonly-used algorithms by applying them to an experimental data set containing small-angle coherent scattering with visibility levels ranging from below 1% to ∼60%. We also propose a contrast calibration protocol and show that a computationally lightweight algorithm can be implemented for high-speed correlation evaluation.
Funder
United States Department of Energy
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献