Abstract
Although simulation models are heavily used in biomechanical research and testing of TKA implants, pre-clinical tools for a holistic estimation of implant performance under dynamic loading conditions are rare. The objective of this study was the development of an efficient pre-clinical test method for analyzing knee contact mechanics and kinematics based on a dynamic FE model and to evaluate the effects of the proximo-distal patella position and the patellar tendon stiffness on the patellar kinematics. A finite element-based workflow for knee prostheses designs was developed based on standardized in vivo load data, which included the tibial forces and moments. In a new research approach, the tibial forces are used as input for the model, whereas the tibial moments were used to validate the results. For the standardized sit down, stand up, and knee bend load cycles, the calculated tibial moments show only small deviations from the reference values—especially for high flexion angles. For the knee bend cycle, the maximum absolute value of patellar flexion decreases for higher patellar tendon stiffness and more distally placed patellar components. Therefore, patella-related clinical problems caused by patella baja may also arise if the patellar tendon is too weak for high tibiofemoral flexion angles.
Subject
Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献