Silicon 3D Microdosimeters for Advanced Quality Assurance in Particle Therapy

Author:

Tran Linh T.ORCID,Bolst David,James Benjamin,Pan Vladimir,Vohradsky James,Peracchi Stefania,Chartier Lachlan,Debrot Emily,Guatelli Susana,Petasecca Marco,Lerch MichaelORCID,Prokopovich Dale,Pastuović Željko,Povoli Marco,Kok Angela,Inaniwa Taku,Lee Sung Hyun,Matsufuji Naruhiro,Rosenfeld Anatoly B.

Abstract

The Centre for Medical Radiation Physics introduced the concept of Silicon On Insulator (SOI) microdosimeters with 3-Dimensional (3D) cylindrical sensitive volumes (SVs) mimicking the dimensions of cells in an array. Several designs of high-definition 3D SVs fabricated using 3D MEMS technology were implemented. 3D SVs were fabricated in different sizes and configurations with diameters between 18 and 30 µm, thicknesses of 2–50 µm and at a pitch of 50 µm in matrices with volumes of 20 × 20 and 50 × 50. SVs were segmented into sub-arrays to reduce capacitance and avoid pile up in high-dose rate pencil beam scanning applications. Detailed TCAD simulations and charge collection studies in individual SVs have been performed. The microdosimetry probe (MicroPlus) is composed of the silicon microdosimeter and low-noise front–end readout electronics housed in a PMMA waterproof sheath that allows measurements of lineal energies as low as 0.4 keV/µm in water or PMMA. Microdosimetric quantities measured with SOI microdosimeters and the MicroPlus probe were used to evaluate the relative biological effectiveness (RBE) of heavy ions and protons delivered by pencil-beam scanning and passive scattering systems in different particle therapy centres. The 3D detectors and MicroPlus probe developed for microdosimetry have the potential to provide confidence in the delivery of RBE optimized particle therapy when introduced into routine clinical practice.

Funder

Australian Research Council

European Space Agency

Publisher

MDPI AG

Subject

Fluid Flow and Transfer Processes,Computer Science Applications,Process Chemistry and Technology,General Engineering,Instrumentation,General Materials Science

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Status of LET assessment with active and passive detectors in ion beams;Radiation Measurements;2024-09

2. Diamond energy-dispersive dosimeter for measuring linear energy deposition distributions in clinical carbon beam therapy;Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms;2024-09

3. LET Calibration of Ion Microbeams and Their SEE Cross Section Characterization;IEEE Transactions on Nuclear Science;2024-08

4. On the radiation quality characterization in radiation therapy: from linear energy transfer to experimental microdosimetry;The European Physical Journal Plus;2024-07-17

5. LETd Optimization Verification With an SOI Microdosimeter;International Journal of Radiation Oncology*Biology*Physics;2024-07