Reverse Torque Value of Angulated Screw Channel Abutment before and after Cyclic Loading: An In Vitro Study

Author:

Chen Yu-Hsuan1,Wu Yu-Ling1,Chen Hung-Shyong23,Lin Ching-Ping1,Wu Aaron Yu-Jen1

Affiliation:

1. Department of Dentistry, Kaohsiung Chang Gung Memorial Hospital and Chang Gung University College of Medicine, Kaohsiung 833, Taiwan

2. Department of Mechanical Engineering, Cheng Shiu University, Kaohsiung 833, Taiwan

3. Center for Environmental Toxin and Emerging-Contaminant Research, Cheng Shiu University, Kaohsiung 833, Taiwan

Abstract

This in vitro experiment aimed to understand the difference in preload acting on an abutment screw under different angles of angulated screw-retained crown and the performance after cyclic loading. In total, thirty implants with angulated screw channel (ASC) abutments were divided into two parts. The first part consisted of three groups: a 0° access channel with a zirconia crown (ASC-0) (n = 5), a 15° access channel with a specially designed zirconia crown (sASC-15) (n = 5), and a 25° access channel with a specially designed zirconia crown (sASC-25) (n = 5). The reverse torque value (RTV) was measured at 0° for each specimen. The second part consisted of three groups: a 0° access channel with a zirconia crown (ASC-0) (n = 5); a 15° access channel with a zirconia crown (ASC-15) (n = 5), and a 25° access channel with a zirconia crown (ASC-25) (n = 5). The manufacturer’s recommended torque was applied to each specimen, and baseline RTV was measured before cyclic loading. Each ASC implant assembly was cyclically loaded at 0 to 40 N with 1 million cycles at 10 Hz. RTV was measured after cyclic loading. Kruskal–Wallis test and Jonckheere–Terpstra test were used for statistical analysis. All specimens were examined under a digital microscope and scanning electron microscope (SEM) to observe the wear of the screw head before and after the whole experiment. A significant difference in the different percentages of straight RTV (sRTV) between the three groups was found (p = 0.027). The angle of ASC to the different percentages of sRTV showed a significant linear trend (p = 0.003). No significant differences were found in RTV difference after cyclic loading among the ASC-0, ASC-15, and ASC-25 groups (p = 0.212). The ASC-25 group had the most serious degree of wear based on a digital microscope and SEM examination. The ASC angle will affect the actual preload acting on a screw: the larger the ASC angle, the smaller the preload. The performance of the angled ASC groups in RTV difference was comparable to that of 0° ASC after cyclic loading.

Funder

Chang Gung Medical Foundation and Cheng Shiu University

Publisher

MDPI AG

Subject

Biomedical Engineering,Biomaterials