Affiliation:
1. Interventional Oncology-Microinvasive Therapy (SIP), Department of Radiology, Medical University Innsbruck, Anichstr. 35, 6020 Innsbruck, Austria
Abstract
Objectives: The purpose of this study was to assess the performance of an optically tracked robot for computed-tomography (CT)-guided needle placements in a phantom study. Methods: In total, 240 needle punctures were carried out with the help of an optically tracked robotic device (Micromate) based on CT image datasets at three different slice thicknesses (1, 3, and 5 mm). Conically shaped targets inside a gelatin-filled plexiglass phantom were punctured. The target positioning error between the planned and actual needle trajectory was assessed by measuring the lateral positioning error (ND) between the target and the puncture needle and the Euclidean distance (ED) between the needle tip and target in control CTs. Results: The mean ND and ED for the thinnest CT slice thickness were 1.34 mm (SD ± 0.82) and 2.1 mm (SD ± 0.75), respectively. There was no significant impact of target depth on targeting accuracy for ND (p = 0.094) or ED (p = 0.187). The mean duration for the planning of one trajectory and for needle positioning were 42 s (SD ± 4) and 64 s (SD ± 7), respectively. Conclusions: In this ex vivo study, the robotic targeting device yielded satisfactory accuracy results at CT slice thicknesses of 1 and 3 mm. This technology may be particularly useful in interventions where the accurate placement of needle-like instruments is required.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献