Affiliation:
1. Department of Nuclear Medicine, Herlev University Hospital, 2730 Herlev, Denmark
Abstract
Three-dimensional echocardiography (3DE) is currently the preferred method for monitoring left ventricular ejection fraction (LVEF) in cancer patients receiving potentially cardiotoxic anti-neoplastic therapy. In Denmark, however, the traditional standard for LVEF monitoring has been rooted in nuclear medicine departments utilizing equilibrium radionuclide angiography (ERNA). Although ERNA remains a principal modality, there is an emerging trend towards the adoption of echocardiography for this purpose. Given this context, assessing the reproducibility of 3DE among non-specialized medical personnel is crucial for its clinical adoption in such departments. To assess the feasibility of 3DE for LVEF measurements by technologists, we evaluated the repeatability and reproducibility of two moderately experienced technologists. They performed 3DE on 12 volunteers over two sessions, with a collaborative review of the results from the first session before the second session. Two-way intraclass correlation values increased from 0.03 to 0.77 across the sessions. This increase in agreement was mainly due to the recognition of false low measurements. Our findings underscore the importance of incorporating reproducibility exercises in the context of 3DE, especially when operated by technologists. Additionally, routine control of the acquisitions by physicians is deemed necessary. Ensuring these hurdles are adequately managed enables the adoption of 3DE for LVEF measurements by technologists.