How Does the Corrected Exhalyzer Software Change the Predictive Value of LCI in Pulmonary Exacerbations in Children with Cystic Fibrosis?

Author:

Wojsyk-Banaszak Irena1ORCID,Stachowiak Zuzanna2ORCID,Więckowska Barbara3ORCID,Andrzejewska Marta1ORCID,Tąpolska-Jóźwiak Katarzyna1ORCID,Szczepankiewicz Aleksandra2ORCID,Sobkowiak Paulina1ORCID,Bręborowicz Anna1

Affiliation:

1. Department of Paediatric Pulmonology, Allergy and Clinical Immunology, Poznan University of Medical Sciences, 60-572 Poznan, Poland

2. Molecular and Cell Biology Unit, Department of Paediatric Pulmonology, Allergy and Clinical Immunology, Poznan University of Medical Sciences, 60-572 Poznan, Poland

3. Department of Computer Science and Statistics, Poznan University of Medical Sciences, 60-806 Poznan, Poland

Abstract

Aim: Recently, the most commonly used for multiple breath washout device, the Exhalyzer D, has been shown to overestimate lung clearance index (LCI) results due to a software error. Our study aimed to compare the predictive values of LCI in the CF pulmonary exacerbations (PE) calculated with the updated (3.3.1) and the previous (3.2.1) version of the Spiroware software. Materials and Methods: The measurements were performed during 259 visits in CF pediatric patients. We used 39ΔPE pairs (PE preceded by stable visit) and 138ΔS pairs (stable visit preceded by stable visit) to compare the LCI changes during PE. The areas under the receiver operating curves (AUCROC) and odds ratios were calculated based on the differences between ΔPEs and ΔSs. The exacerbation risk was estimated using a logistic regression model with generalized estimating equations (GEE). Results: There were statistically significant differences in LCI 2.5% median values measured using the two versions of the software in the stable condition but not during PE. The AUCROC for changes between the two consecutive visits for LCI did not change significantly using the updated Spiroware software. Conclusions: Despite the lower median values, using the recalculated LCI values does not influence the diagnostic accuracy of this parameter in CF PE.

Funder

Polish National Science Centre

Publisher

MDPI AG

Subject

Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3