Optimization of Kernel Type and Sharpness Level Improves Objective and Subjective Image Quality for High-Pitch Photon Counting Coronary CT Angiography

Author:

Yang Yang1,Fink Nicola23,Emrich Tilman12ORCID,Graafen Dirk1,Richter Rosa1,Bockius Stefanie1,Wolf Elias V.1ORCID,Laux Gerald4,Kavermann Larissa4,Müller Lukas1,Hell Michaela4,Halfmann Moritz C.1ORCID

Affiliation:

1. Department of Diagnostic and Interventional Radiology, University Medical Center Mainz, Langenbeckstrasse 1, 55131 Mainz, Germany

2. Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC 29425, USA

3. Department of Radiology, University Hospital, LMU Munich, 81377 Munich, Germany

4. Department of Cardiology, University Medical Center Mainz, Langenbeckstrasse 1, 55131 Mainz, Germany

Abstract

(1) Background: Photon-counting detector (PCD) CT offers a wide variety of kernels and sharpness levels for image reconstruction. The aim of this retrospective study was to determine optimal settings for coronary CT angiography (CCTA). (2) Methods: Thirty patients (eight female, mean age 63 ± 13 years) underwent PCD-CCTA in a high-pitch mode. Images were reconstructed using three different kernels and four sharpness levels (Br36/40/44/48, Bv36/40/44/48, and Qr36/40/44/48). To analyze objective image quality, the attenuation, image noise, contrast-to-noise ratio (CNR), and vessel sharpness were quantified in proximal and distal coronaries. For subjective image quality, two blinded readers assessed image noise, visually sharp reproduction of coronaries, and the overall image quality using a five-point Likert scale. (3) Results: Attenuation, image noise, CNR, and vessel sharpness significantly differed across kernels (all p < 0.001), with the Br-kernel reaching the highest attenuation. With increasing kernel sharpness, image noise and vessel sharpness increased, whereas CNR continuously decreased. Reconstruction with Br-kernel generally had the highest CNR (Br > Bv > Qr), except Bv-kernel had a superior CNR at sharpness level 40. Bv-kernel had significantly higher vessel sharpness than Br- and Qr-kernel (p < 0.001). Subjective image quality was rated best for kernels Bv40 and Bv36, followed by Br36 and Qr36. (4) Conclusion: Reconstructions with kernel Bv40 are beneficial to achieve optimal image quality in spectral high-pitch CCTA using PCD-CT.

Publisher

MDPI AG

Subject

Clinical Biochemistry

Reference34 articles.

1. A Review on Coronary Artery Disease, Its Risk Factors, and Therapeutics;Malakar;J. Cell. Physiol.,2019

2. SCOT-HEART Investigators, Newby, D.E., Adamson, P.D., Berry, C., Boon, N.A., Dweck, M.R., Flather, M., Forbes, J., Hunter, A., and Lewis, S. (2018). Coronary Ct Angiography and 5-Year Risk of Myocardial Infarction. N. Engl. J. Med., 379, 924–933.

3. 2021 Aha/Acc/Ase/Chest/Saem/Scct/Scmr Guideline for the Evaluation and Diagnosis of Chest Pain: Executive Summary: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines;Gulati;Circulation,2021

4. 2019 Esc Guidelines for the Diagnosis and Management of Chronic Coronary Syndromes;Knuuti;Eur. Heart J.,2020

5. 2019 Guidelines for the Diagnosis and Management of Chronic Coronary Syndromes: Recommendations for Cardiovascular Imaging;Saraste;Herz,2020

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3