A Finite Element Analysis Study from 3D CT to Predict Transcatheter Heart Valve Thrombosis

Author:

Nappi Francesco,Mazzocchi Laura,Timofeva Irina,Macron Laurent,Morganti SimoneORCID,Avtaar Singh Sanjeet Singh,Attias David,Congedo Antonio,Auricchio Ferdinando

Abstract

Background: Transcatheter aortic valve replacement has proved its safety and effectiveness in intermediate- to high-risk and inoperable patients with severe aortic stenosis. However, despite current guideline recommendations, the use of transcatheter aortic valve replacement (TAVR) to treat severe aortic valve stenosis caused by degenerative leaflet thickening and calcification has not been widely adopted in low-risk patients. This reluctance among both cardiac surgeons and cardiologists could be due to concerns regarding clinical and subclinical valve thrombosis. Stent performance alongside increased aortic root and leaflet stresses in surgical bioprostheses has been correlated with complications such as thrombosis, migration and structural valve degeneration. Materials and Methods: Self-expandable catheter-based aortic valve replacement (Medtronic, Minneapolis, MN, USA), which was received by patients who developed transcatheter heart valve thrombosis, was investigated using high-resolution biomodelling from computed tomography scanning. Calcific blocks were extracted from a 250 CT multi-slice image for precise three-dimensional geometry image reconstruction of the root and leaflets. Results: Distortion of the stent was observed with incomplete cranial and caudal expansion of the device. The incomplete deployment of the stent was evident in the presence of uncrushed refractory bulky calcifications. This resulted in incomplete alignment of the device within the aortic root and potential dislodgment. Conclusion: A Finite Element Analysis (FEA) investigation can anticipate the presence of calcified refractory blocks, the deformation of the prosthetic stent and the development of paravalvular orifice, and it may prevent subclinical and clinical TAVR thrombosis. Here we clearly demonstrate that using exact geometry from high-resolution CT scans in association with FEA allows detection of persistent bulky calcifications that may contribute to thrombus formation after TAVR procedure.

Publisher

MDPI AG

Subject

Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3