Abstract
Background: Point-of-care testing (POCT) provides shorter turn-around times and, in many cases, potentially improves medical decision making. The AQT90 FLEX® benchtop immunoanalyzer (Radiometer Medical ApS, Copenhagen, Denmark) allows for the determination of beta-human chorionic gonadotropin (βhCG) in 18 min. The main aim of this study was to evaluate the impact of measuring βhCG using the AQT90 analyzer in the gynecology emergency department (ED) compared to the standard practice of using central laboratory blood testing on the patient length of stay (LOS). Methods: The evaluation consisted of two parts. The first one, conducted in the central laboratory, focused on the analytical performances of the AQT βhCG assay. The second one, conducted in the ED, aimed at determining the impact of POCT βhCG implementation on the timeframe in which ED patients require βhCG assessment. Results: The within-lab imprecisions at the mean values of 17 and 287 IU/L were 2.7% and 3.7%, respectively. Using Deming regression (n = 60), the following equation was obtained in the central lab: AQT90 βhCG = 1.1 Roche βhCG—12.9 (r = 0.997). The implementation of POCT βhCG in the ED significantly reduced patient LOS (145 (90–212) min vs. 205 (155–265) with and without AQT90, respectively, p < 0.001). At the 2 IU/L decision level, a 99.7% agreement with the Roche assay was reported (kappa statistics, 0.99). Conclusions: We confirm that the analytical qualities of the AQT 90 were in line with those obtained in the central lab. The implementation of the POCT βhCG is associated with a shorter LOS in the ED due to the faster availability of the results and the faster decision-making possibilities.