An Evaluation of the Intubrite Laryngoscope in Simulated In-Hospital and Out-of-Hospital Settings by Individuals with No Clinical Experience: A Randomized, Cross-Over, Manikin Study

Author:

Ratajczyk Paweł,Fedorczak MichałORCID,Gaszyński TomaszORCID

Abstract

Introduction: The aim of the study was to compare the Intubrite laryngoscope and the standard Macintosh blade laryngoscope (MCL) used by persons with no clinical experience in simulated hospital and non-hospital conditions on a manikin model. Materials and methods: The study involved 50 students of Medical Rescue. The hospital conditions (intubation height 110 cm—high position) and those occurring at the scene (intubation of a manikin located at floor level-low position) were simulated. The analysis included: duration of intubation, visibility of the laryngeal opening by the Cormack–Lehane scale, the bioelectrical activity of the intubating muscles, comfort and subjective assessment of physical effort by the Borg scale (Borg’s scale of subjective feeling of effort). The statistical analysis was performed with Microsoft Excel and T-student tests for pairs with unequal variables. The statistical importance was set at p < 0.05. Results: The use of an Intubrite laryngoscope significantly reduces the mean endotracheal intubation time compared to the Macintosh laryngoscope in a low position (17.34 s versus 19.04 s, p < 0.05). A higher rate of repeat intubations was observed in a low position for MCL (from 10% to 14%, p < 0.05). The reverse is true for Intubrite laryngoscope intubation (from 8% to 4%, p < 0.05 Please explain what is compared). The Intubrite laryngoscope improved visualization of glottis in the high and the low positions compared to the Macintosh laryngoscope (54% and 50% to 52% and 38%, respectively, p < 0.05). The risk of tooth damage was the same for the Intubrite and the MCL laryngoscopes in a high position (16% and 14%, respectively, p > 0.05), while in the simulated out-of-hospital setting, it was significantly higher for the MCL (22% versus 8%, p < 0.05). The subjective comfort of intubation in both simulated situations was similar according to the Borg scale (p-value, values). The use of the Intubrite laryngoscope was associated with less effort than the MCL in high versus low positions. For MCL, intubation in a high position was associated with lower mean muscle activity than in a low position (+48.24 µV versus +58.25 µV, p-value). For the Intubrite laryngoscope, these values were at similar levels (+52.03 µV and +52.65 µV, p-value). Conclusions: The use of the Intubrite laryngoscope by people with no clinical experience shortens the time of intubation and improves the laryngeal view compared to the standard Macintosh laryngoscope, but it requires similar muscle work in simulated conditions.

Publisher

MDPI AG

Subject

Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3