Impact of Phantom Size on Low-Energy Virtual Monoenergetic Images of Three Dual-Energy CT Platforms

Author:

Greffier Joël1ORCID,Van Ngoc Ty Claire2ORCID,Fitton Isabelle2ORCID,Frandon Julien1ORCID,Beregi Jean-Paul1,Dabli Djamel1ORCID

Affiliation:

1. IMAGINE UR UM 103, Department of Medical Imaging, Nimes University Hospital, Montpellier University, 30029 Nimes, France

2. Department of Radiology, Assistance Publique Hôpitaux de Paris, Hôpital Européen Georges Pompidou, Université de Paris, 75015 Paris, France

Abstract

The purpose of this study was to compare the quality of low-energy virtual monoenergetic images (VMIs) obtained with three Dual-Energy CT (DECT) platforms according to the phantom diameter. Three sections of the Mercury Phantom 4.0 were scanned on two generations of split-filter CTs (SFCT-1st and SFCT-2nd) and on one Dual-source CT (DSCT). The noise power spectrum (NPS), task-based transfer function (TTF), and detectability index (d’) were assessed on VMIs from 40 to 70 keV. The highest noise magnitude values were found with SFCT-1st and noise magnitude was higher with DSCT than with SFCT-2nd for 26 cm (10.2% ± 1.3%) and 31 cm (7.0% ± 2.5%), and the opposite for 36 cm (−4.2% ± 2.5%). The highest average NPS spatial frequencies and TTF values at 50% (f50) values were found with DSCT. For all energy levels, the f50 values were higher with SFCT-2nd than SFCT-1st for 26 cm (3.2% ± 0.4%) and the opposite for 31 cm (−6.9% ± 0.5%) and 36 cm (−5.6% ± 0.7%). The lowest d’ values were found with SFCT-1st. For all energy levels, the d’ values were lower with DSCT than with SFCT-2nd for 26 cm (−6.2% ± 0.7%), similar for 31 cm (−0.3% ± 1.9%) and higher for 36 cm (5.4% ± 2.7%). In conclusion, compared to SFCT-1st, SFCT-2nd exhibited a lower noise magnitude and higher detectability values. Compared with DSCT, SFCT-2nd had a lower noise magnitude and higher detectability for the 26 cm, but the opposite was true for the 36 cm.

Publisher

MDPI AG

Subject

Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3