Smartphone Slit Lamp Imaging—Usability and Quality Assessment

Author:

Muth Daniel Rudolf1ORCID,Blaser Frank1,Foa Nastasia1,Scherm Pauline2,Mayer Wolfgang Johann2,Barthelmes Daniel1,Zweifel Sandrine Anne1ORCID

Affiliation:

1. Department of Ophthalmology, University Hospital Zurich, University of Zurich, Frauenklinikstrasse 24, 8091 Zurich, Switzerland

2. Department of Ophthalmology, University Hospital, LMU Munich, Mathildenstrasse 8, 80336 Munich, Germany

Abstract

Purpose: To assess the usability and image quality of a smartphone adapter for direct slit lamp imaging. Methods: A single-center, prospective, clinical study conducted in the Department of Ophthalmology at the University Hospital Zurich, Switzerland. The smartphone group consisted of 26 medical staff (consultants, residents, and students). The control group consisted of one ophthalmic photographer. Both groups took images of the anterior and the posterior eye segment of the same proband. The control group used professional photography equipment. The participant group used an Apple iPhone 11 mounted on a slit lamp via a removable SlitREC smartphone adapter (Custom Surgical GmbH, Munich, Germany). The image quality was graded independently by two blinded ophthalmologists on a scale from 0 (low) to 10 (high quality). Images with a score ≥ 7.0/10 were considered as good as the reference images. The acquisition time was measured. A questionnaire on usability and experience in smartphone and slit lamp use was taken by all of the participants. Results: Each participant had three attempts at the same task. The overall smartphone quality was 7.2/10 for the anterior and 6.4/10 for the posterior segment. The subjectively perceived difficulty decreased significantly over the course of three attempts (Kendall’s W). Image quality increased as well but did not improve significantly from take 1 to take 3. However, the image quality of the posterior segment was significantly, positively correlated (Spearman’s Rho) with work experience. The mean acquisition time for anterior segment imaging was faster in the smartphone group compared to the control group (156 vs. 206 s). It was vice versa for the posterior segment (180 vs. 151 s). Conclusion: Slit lamp imaging with the presented smartphone adapter provides high-quality imaging of the anterior segment. Posterior segment imaging remains challenging in terms of image quality. The adapter constitutes a cost-effective, portable, easy-to-use solution for recording ophthalmic photos and videos. It can facilitate clinical documentation and communication among colleagues and with the patient especially outside normal consultation hours. Direct slit lamp imaging allows for time to be saved and increases the independence of ophthalmologists in terms of patient mobility and the availability of photographic staff.

Publisher

MDPI AG

Subject

Clinical Biochemistry

Reference34 articles.

1. The ophthalmoscope: Helmholtz’s Augenspiegel;Pearce;Eur. Neurol.,2009

2. Helmholtz, H. (1851). Beschreibung Eines Augen-Spiegels zur Untersuchung der Netzhaut im Lebenden Auge, Springer.

3. Gellrich, M.-M. (2014). The Slit Lamp, Springer.

4. A new view of the slit lamp;Gellrich;Br. J. Ophthalmol.,2009

5. Kalayoglu, M.V. (2022, October 04). The Evolution of Slit Lamp Biomicroscopy. Available online: https://www.ophthalmologyweb.com/Tech-Spotlights/26450-The-Evolution-of-Slit-Lamp-Biomicroscopy/.

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3